Model
CA20
Thermocouple Calibrator
Accepts Type J Thermocouples, High/Low Alarm, Selectable Resolution.

 
Dual function CA20 Thermocouple Calibrator can be used to source Type J thermocouple output values or measure Type J thermocouples. Quickly select output value using the rotary switch with coarse and fine adjustments. Model CA20 accepts up to two Type J thermocouples and features data hold and max functions. Measurements are displayed in a 3½-digit backlit display. Units include a 4" (100 mm) Type J bead wire temperature probe, calibration cable, 9 V battery and instruction manual.


 

PHYSICAL DATA
Thermocouple Type: Type J.
Measurement Range: -40 to 1400°F (-40 to 760°C), selectable.
Calibration Range: -328 to 2192°F (-210 to 1200°C).
Accuracy: ±0.1% rdg + 2°F on -40 to 1400°F; ±0.1% rdg +1°C on -40 to 760°C, over operating temperature of 64 to 82°F (18 to 28°C).
Resolution: Selectable 1° or 0.1 °.
Temperature Coefficient: 0.1 times the applicable accuracy specification per °F from 32 to 64°F and 82 to 122°F (0 to 18°C and 28 to 50°C).
Reading Rate: 2.5 times/second.
Input Protection: 24 VDC or 24 VAC rms max input voltage on any combination of input pins.
Operating Temperature: 32 to 122°F (0 to 50°C) max 80% RH.
Storage Temperature: -4 to 140°F (-20 of 60°C) max 70% RH.
Display: 0.8" (20 mm) height, 3½-digit LCD with switchable back light.
Thermocouple Connection: Standard female mini-connector.
Battery: 9V Alkaline (Included).
Battery Life: 200 hours typical.
Housing: ABS Plastic.
Weight: 15.5 oz. (440 g).
Applied Thermal Systems, Inc. © 2010 | All Rights reserved